Products

Atomic Absorption, Fully automatic flame and furnace type

Integrated flame/graphite furnace 
Wavelength range 190-900nm
Wavelength accuracy Better than ±0.25nm
Spectral bandwidth 0.1 nm, 0.2nm, 0.4nm, 1.2nm, auto switch overGrating 1800 l/mm 
Focal length 277mm 
Blazed wavelength 250nm
Adopting FUZZY-PID and dual curve mode
light-controlled temperature control technique
The temperature control accuracy is less than 1%
Easy and practical analysis software
● Adopting large-scale programmable logic array & 
Inter I2C bus tech
● Multi-function autosampler
Characteristic Mass of Some Elements using rich
oxygen air-C2H2 flame and other flame methods
● Automatic wavelength scanning and peak picking. 
● Automatic spectral bandwidth changing.
Autosampler 55 sample vessels


Integrated flame/graphite furnace 
Wavelength range 190-900nm
Wavelength accuracy Better than ±0.25nm
Spectral bandwidth 0.1 nm, 0.2nm, 0.4nm, 1.2nm, auto switch overGrating 1800 l/mm 
Focal length 277mm 
Blazed wavelength 250nm
Adopting FUZZY-PID and dual curve mode
light-controlled temperature control technique
The temperature control accuracy is less than 1%
Easy and practical analysis software
● Adopting large-scale programmable logic array & 
Inter I2C bus tech
● Multi-function autosampler
Characteristic Mass of Some Elements using rich
oxygen air-C2H2 flame and other flame methods
● Automatic wavelength scanning and peak picking. 
● Automatic spectral bandwidth changing.
Autosampler 55 sample vessels



SPECIFICATIONS

 

 

 

 

Main Specification

 

 

 

 

 

Wavelength range

190-900nm

Wavelength accuracy

Better than ±0.25nm

Resolution

Two spectral lines of Mn at 279.5nm and 279.8nm can be

separated with the spectral bandwidth of 0.2nm and

valley-peak energy ratio less than 30%.

Baseline stability

0.004A/30min

Background correction

The D2 lamp background correction capability at 1A is better than 30 times.

The S-H background correction capability at 1.8A is better than 30 times.

Light Source System

Lamp turret

Motorized 6-lamp turret (Two high performance HCLs can be

mounted on the turret to increase the sensitivity in flame analysis.)

Lamp current adjustment

Wide pulse current: 0~25mA,

Narrow pulse current: 0~10mA.

Lamp power supply mode

400Hz square wave pulse; 100Hz narrow square wave pulse +

400Hz wide square wave pulse.

Optical System

Monochromator

Single beam, Czerny-Turner design grating monochromator

Grating

1800 l/mm

Focal length

277mm

Blazed wavelength

250nm

Spectral bandwidth

0.1 nm, 0.2nm, 0.4nm, 1.2nm, auto switch over

Flame Atomizer

Burner

10cm single slot all-titanium burner

Spray chamber

Corrosion resistant all-plastic spray chamber.

Nebulizer

High efficiency glass nebulizer with metal sleeve.

sucking up rate: 6-7mL/min

Emission burner provided

Graphite Furnace

Temperature range

Room temperature-3000°C

Heating rate

2000°C/s

Graphite tube dimensions

28mm (L) x 8mm (OD)

Characteristic mass

Cd?0.8*10-12g, Cu?5*10-12g, Mo?1*10-11g

Precision

Cd?3%, Cu?3%, Mo?4%

Detection and Data

Processing System

Detector

R928 photomultiplier with high sensitivity and wide

spectral range.

Software

Under Windows operating system

Analytical method

Working curve auto-fitting; standard addition method;

automatic sensitivity correction;

automatic calculation of concentration and content.

Repeat times

1~99 times. automatic calculation of mean value, standard

deviation and relative standard deviation

Multi-task function

Sequential determination of multi-elements in the same sample.

Condition reading

With model function

Result printing

Measurement data and final analytical report printout, editing

with Excel.

Standard RS-232 serial port communication

Graphite Furnace

Autosampler

Sample tray capacity

55 sample vessels and 5 reagent vessels

Vessel material

Polypropylene

Vessel volume

3ml for sample vessel. 20ml for reagent vessel

Minimum sampling volume

1µl

Repeatable sampling times

1~99 times

Sampling system

Accurate dual pump system. with 100µI and 1 ml injectors.

Characteristic

Concentration &

Detection Limit

Air-C2H2 flame

Rich oxygen Air-C2H2 flame

Cu: Characteristic concentration ? 0.025 mg/L,

Detection limit?0.006mg/L;

Ba: Characteristic concentration ? 0.22mg/L

AI: Characteristic concentration ? 0.4mg/L

Function Expansion

Hydridee vapor generator can be connected for Hydridee analysis.

Dimensions and Weight

Main unit

107X49x58cm,140kg

Graphite furnace

42X42X46cm, 65kg

Autosampler

40X29X29cm, 15kg